Engineering Toolbox/Calibration Error

Instrumentation & Calibration

Calibration Error

Compare an instrument indication against an applied reference and express the signed error in engineering units and percentage of calibrated span.

Calibration point

Error = indicated value − applied/reference value.

% span

Field use

One calibration point, clearly assessed.

Use this when checking a single applied point against the instrument indication. The tolerance entered here must come from the approved calibration requirement or applicable procedure.

Calculation basis

Error = Indicated − Applied
% span error = (Error / |URV − LRV|) × 100

A negative error means the instrument indicates below the applied reference. A positive error means it indicates above the applied reference.

Pass/fail compares the absolute % span error against the entered tolerance. AccessRA does not assume an instrument accuracy class or site calibration tolerance.

Interpretation

What the sign tells you

A positive error means the instrument reads higher than the applied reference; a negative error means it reads lower. The sign helps describe direction, while pass/fail uses the absolute magnitude against the entered tolerance.

Worked example

0–1000 psi gauge at 500 psi

If the reference is 500 psi and the instrument indicates 515 psi, the signed error is +15 psi. With a 1000 psi calibrated span, the error is +1.5% of span. A ±1% span tolerance would therefore fail at that point.

Practical boundary

One point does not describe the instrument

A single-point check cannot by itself distinguish zero, span, linearity or hysteresis behaviour. Use a multi-point upscale/downscale calibration when the condition of the complete measurement range must be assessed.